called capture.txt and the file will be an ASCII text
file.
11. You should see the data on the computer screen.
12. When you are finished storing data from the
Nanopure, save your file. You can exit Hyperterminal
or set up a new experiment.
13. Hyperterminal data can be manipulated and
graphed in Excel
®
.
Procomm
1. Connect the 9 pin serial cable from the Nanopure
RS-232 port to COM2: port or (COM1: port) on
back of computer.
2. From the (Idle) display, press the UP or DOWN
arrow until the display reads, “Advanced Menu.”
3. Press ENTER.
4. From the “Printer Attached” display. Press the Up or
DOWN arrow for (NO).
5. Press ENTER.
6. The unit will now, during the normal Recirculation
mode send a purity and temperature reading to the
computer once per minute. If “Use TOC Analyzer”
has been set to “Yes,” the TOC data will also be
sent.
7. Open DOS window and change directory to pro-
comm directory.
8. Open procomm executable.
9. Press ALT-F10 for configuration screen.
10. Enter into Line Settings screen by typing ALT-P.
11. Type in 11 <Enter>, sets parameters as 9600, 8, N,
1.
COMPUTER/PRINTER SETUP