7. Pull the ultrafilter out of the unit and turn it to the
left side. (This will make it easier to disconnect
the bottom fitting.)
8. Using the tool supplied with the ultrafilter, simul-
taneously push and hold the collet on the ultrafil-
ter in while pulling the bottom fitting out of the
connector. See photo.
9. Install new ultrafilter according to “Ultrafilter
Installation” in the “Installation” section of this
manual.
10. Clean the Nanopure TOC - UF according to the
“System Cleaning” section of this manual.
MAINTENANCE AND SERVICING
Collet
Figure 19: Tool used for ultrafilter removal
Flow
Figure 21: Ultrafilter Connections - Right side view
Tubing with
luer fitting
Metal bracket
Velcro strap
Bottom connector
Top connector
Figure 20: Ultrafilter Installation